Продукция

CP030-3M

CP030-3M

30 A, 10 MHz Current Probe - AC/DC, 30 A rms, 50 A Peak Pulse, 3 meter cable
T3CP50 Токовый пробник для осциллографов. Полоса пропускания 50 МГц. Время нарастания ≤7 нс. 2 диапазона: 100 мВ/А, 1 В/А. Максимальный ток: 50 Aпик-пик или 30 Аскз. Погрешность ±1 %. ф
DCS025 Deskew Calibration Source for CP030, CP030A, CP031, CP031A, AP015, CP150, CP500
CP150-6M 150 A, 5 MHz Current Probe - AC/DC, 150 A rms, 500 A Peak Pulse, 6 meter cable
CP030-3M 30 A, 10 MHz Current Probe - AC/DC, 30 A rms, 50 A Peak Pulse, 3 meter cable
DCS015 Deskew Calibration Source for AP015
CA10-QUADPAK Programmable ProBus Current Adapter
CA10 Programmable ProBus Current Adapter
CP500 Токовый пробник до 500 А, 2 МГц
CP150 Токовый пробник до 150 А, максимум 500 A < 30 мкс, 10 МГц
CP031A Высокочувствительный токовый пробник до 30 A, 100 МГц
CP031 Токовый пробник до 30 A, 100 МГц
CP030A Высокочувствительный токовый пробник до 30 A, 50 МГц
CP030 Токовый пробник до 30 A, 50 МГц

Teledyne LeCroy current probes are available in a variety of models for a wide range of applications. The full range of Teledyne LeCroy current probes includes models with bandwidths up to 100 MHz, peak currents up to 700 A and sensitivities to 1 mA/div. Teledyne LeCroy current probes are often used in applications such as the design and test of switching power supplies, motor drives, electric vehicles, and uninterruptible power supplies.

  • CP030 - 30A, 50MHz Current Probe
  • CP030A - 30A, 50MHz High Sensitivity Current Probe
  • CP031 - 30A, 100MHz Current Probe
  • CP031A - 30A, 100MHz High Sensitivity Current Probe
  • CP150 - 150 A, 10 MHz Current Probe - AC/DC, 150 A rms, 500 A Peak Pulse
  • CP500 - 500 A, 2 MHz Current Probe - AC/DC, 500 A rms, 700 A Peak Pulse
  • DCS015 - Deskew Calibration Source for CP031,CP031A, CP030, CP030A and AP015

Current Sensor Adapter

The CA10 is a programmable and customizable interface device that seamlessly incorporates third party current transducers/transformers with Teledyne LeCroy oscilloscopes or motor drive analyzers. The easy to use interface provides the ability for the CA10 to be programmed to contain the specifications of the current sensor allowing it to automatically correct for the gain or attenuation and display results in Ampere units. This allows the third party device to be recognized and operate as if it were a Teledyne LeCroy probe.

  • CA10 - Programmable ProBus Current Sensor Adapter for use with third party current sensors
  • CA10-QUADPAK - Set of 4 CA10, Programmable ProBus Current Sensor Adapter for use with third party current sensors

Electrical Characteristics

Bandwidth (Probe Only) 10 MHz
Coupling AC 1 MΩ, DC 1MΩ, GND
Risetime (typical) 35 ns
Max Continuous Input Current 30 A
Max Peak Current
(non-continous)
50 A
Minimum Sensitivity 10 mA/div
Amplitude Accuracy ±1% of reading to 30 A
±2% of reading to 50 A
Maximum Circuit Voltage 300 V, CAT I
(Insulated Wire)
 

Environmental and Safety

Operating Temperature 0 °C to 40 °C
Humidity ≤ 80 % relative humidity
(non-condensing) 
Storage Temperature -10 °C to 50 °C
Max Altitude 2000 m
 

Physical Dimensions

Probe Head Dimensions (L x W x H) 175 mm x 18 mm x 40 mm
Cable Length 3 m
Maximum Conductor Size 5 mm
Weight 290 g
 

General Characteristics

Interface ProBus
WaveMaster / SDA 8 Zi-B-R
Узнать больше